This study1 tested the accuracy of the DINAMAPTM SuperSTAT blood pressure technology in neonates and infants according to the ANSI/AAMI SP-10:2002 Standard.
The study took place in neonatal intensive care units and cardiac cath labs. Systolic, diastolic, mean arterial pressure (MAP), and pulse rate (PR) were compared to intra-arterial blood pressure (IBP) measurements. The differences between the SuperSTAT algorithm measurements and the IBP reference were analyzed.
The SuperSTAT NIBP algorithm met the ANSI/AAMI SP-10:2002 accuracy specifications of ±5 mmHg for mean difference and ≤ 8 mmHg standard deviation for the neonatal/infant population.
1 Diego Lozano, MD., Andrea Stebor, RN, Ph.D., Rana F. Mikhail, Ph.D